Controlled environment fumigation chambers for the study of reactive air pollutant effects on plants.

Research output: Contribution to journalJournal article

Published

Journal publication date04/1993
JournalAtmospheric Environment. Part A. General Topics
Journal number5
Volume27
Number of pages6
Pages679-684
Original languageEnglish

Abstract

The design and construction of eight 1-m3 fumigation chambers for exposing plants to reactive air pollutants at low concentrations are reported. Teflon surfaces are used where possible to minimize the adsorption, absorption, desorption and chemical reaction of the pollutants. A purified ambient air supply, to which metered quantities of gaseous pollutants are added, is used to give two air changes per minute at constant, low, pollutant concentrations. Comparative analysis of the chambers indicates that conditions may be maintained with a significant degree of precision, i.e. temperature ±0.3°C, RH ±6%, light intensity ±5 μmol m−2 s−1. Boundary layer analysis from models of cherry tree (Prunus avium) leaves indicate that the minimum conductance value within these chambers is 2 cm s−1.