12,000

We have over 12,000 students, from over 100 countries, within one of the safest campuses in the UK

97%

97% of Lancaster students go into work or further study within six months of graduating

Home > Research > Publications & Outputs > Controlled environment fumigation chambers for ...
View graph of relations

« Back

Controlled environment fumigation chambers for the study of reactive air pollutant effects on plants.

Research output: Contribution to journalJournal article

Published

<mark>Journal publication date</mark>04/1993
<mark>Journal</mark>Atmospheric Environment. Part A. General Topics
Issue5
Volume27
Number of pages6
Pages679-684
<mark>Original language</mark>English

Abstract

The design and construction of eight 1-m3 fumigation chambers for exposing plants to reactive air pollutants at low concentrations are reported. Teflon surfaces are used where possible to minimize the adsorption, absorption, desorption and chemical reaction of the pollutants. A purified ambient air supply, to which metered quantities of gaseous pollutants are added, is used to give two air changes per minute at constant, low, pollutant concentrations. Comparative analysis of the chambers indicates that conditions may be maintained with a significant degree of precision, i.e. temperature ±0.3°C, RH ±6%, light intensity ±5 μmol m−2 s−1. Boundary layer analysis from models of cherry tree (Prunus avium) leaves indicate that the minimum conductance value within these chambers is 2 cm s−1.