A Monte Carlo study of the effect of coded-aperture material and thickness on neutron imaging

Research output: Contribution to journalJournal article

Published

Journal publication date2014
JournalRadiation Protection Dosimetry
Early online date20/11/13
Original languageEnglish

Abstract

In this paper, a coded-aperture design for a scintillator-based neutron imaging system has been simulated using a series of Monte Carlo simulations. Using Monte Carlo simulations, work to optimise a system making use of the EJ-426 neutron scintillator detector has been conducted. This type of scintillator has a low sensitivity to gamma rays and is therefore particularly useful for neutron detection in a mixed radiation environment. Simulations have been conducted using varying coded-aperture materials and different coded-aperture thicknesses. From this, neutron images have been produced, compared qualitatively and quantitatively for each case to find the best material for the MURA (modified uniformly redundant array) pattern. The neutron images generated also allow observations on how differing thicknesses of coded-aperture impact the system.