    Rights statement: ©2018 IEEE. Personal use of this material is permitted. However, permission to reprint/republish this material for advertising or promotional purposes or for creating new collective works for resale or redistribution to servers or lists, or to reuse any copyrighted component of this work in other works must be obtained from the IEEE.

    Accepted author manuscript, 506 KB, PDF document

    Available under license: CC BY-NC: Creative Commons Attribution-NonCommercial 4.0 International License

Links

Text available via DOI:

View graph of relations

Characterisation of a pixelated plastic scintillator for a coded aperture neutron/gamma imaging system

Research output: Contribution in Book/Report/Proceedings - With ISBN/ISSNConference contribution/Paperpeer-review

Published
Publication date5/09/2019
Host publication2018 IEEE Nuclear Science Symposium and Medical Imaging Conference, NSS/MIC 2018 - Proceedings
PublisherIEEE
ISBN (electronic)9781538684948
<mark>Original language</mark>English

Abstract

This paper experimentally investigates, the suitability of an organic pixelated plastic scintillator for a coded aperture neutron/gamma imaging system. The scintillator used in this study has been designed as a small scaled detector with an individual pixel size of 2.8 mm × 2.8 mm × 15 mm. Individual blocks of the scintillator have been separated from one another with ESR TM reflector foil to provide over 70% of optical isolation between pixels. Individual scintillator cells are arranged into 13×13 array with overall dimension of 39.52 mm × 39.52 mm. In this study the scintillator was attached to a single channel photomultiplier tube to assess its pulse shape discrimination capabilities. Initially the scintillator was irradiated with 137Cs gamma-ray source, and gamma-ray pulses were benchmarked with respect to a mathematical model. The detector was then irradiated with a spontaneous fission source, 252 Cf, and preliminary results suggested good pulse shape discrimination potential of the scintillator. It is believed, that it is the first time a small scale pixelated organic plastic scintillator has been tested in context of PSD capabilities for a coded aperture neutron/gamma imaging system.

Bibliographic note

©2018 IEEE. Personal use of this material is permitted. However, permission to reprint/republish this material for advertising or promotional purposes or for creating new collective works for resale or redistribution to servers or lists, or to reuse any copyrighted component of this work in other works must be obtained from the IEEE.