Compact Back-End Electronics with Temperature Compensation and Efficient Data Management for In Situ SiPM-Based Radiation Detection. / Dixon, Nile; Monk, Stephen; Graham, James et al.
In: Sensors, Vol. 23, No. 8, 4053, 17.04.2023.

Research output: Contribution to Journal/MagazineJournal articlepeer-review

Harvard

APA

Vancouver

Author

Bibtex

@article{628eb77a238b4172b555d4660db6f7d4,
title = "Compact Back-End Electronics with Temperature Compensation and Efficient Data Management for In Situ SiPM-Based Radiation Detection",
abstract = "A compact back-end interface for silicon photomultipliers (SiPMs) implementing Zener diode-based temperature compensation has been developed for the remote detection of beta and gamma radiation. Remote detection is facilitated by the development of an efficient data management system utilising MySQL database storage for recording periodic spectra data for wireless access over a private Wi-Fi network. A trapezoidal peak shaping algorithm has been implemented on an FPGA for the continuous conversation of pulses from the SiPM, signifying the detection of a radiological particle, into spectra. This system has been designed to fit within a 46 mm cylindrical diameter for in situ characterization, and can be attached to one or more SiPMs used in conjunction with a range of scintillators. LED blink tests have been used to optimise the trapezoidal shaper coefficients to maximise the resolution of the recorded spectra. Experiments with an array of SiPMs integrated with a NaI(Tl) scintillator exposed to sealed sources of Co-60, Cs-137, Na-22 and Am-241 have shown that the detector achieves a peak efficiency of 27.09 ± 0.13% for a gamma peak at 59.54 keV produced by Am-241, and a minimum energy resolution (Delta E/E) of 4.27 ± 1.16% for the 1332.5 keV gamma peak from Co-60.",
keywords = "silicon photomultiplier, gamma detectors, temperature compensation, wireless data acquisition systems",
author = "Nile Dixon and Stephen Monk and James Graham and David Cheneler",
year = "2023",
month = apr,
day = "17",
doi = "10.3390/s23084053",
language = "English",
volume = "23",
journal = "Sensors",
issn = "1424-8220",
publisher = "Multidisciplinary Digital Publishing Institute (MDPI)",
number = "8",

}

RIS

TY - JOUR

T1 - Compact Back-End Electronics with Temperature Compensation and Efficient Data Management for In Situ SiPM-Based Radiation Detection

AU - Dixon, Nile

AU - Monk, Stephen

AU - Graham, James

AU - Cheneler, David

PY - 2023/4/17

Y1 - 2023/4/17

N2 - A compact back-end interface for silicon photomultipliers (SiPMs) implementing Zener diode-based temperature compensation has been developed for the remote detection of beta and gamma radiation. Remote detection is facilitated by the development of an efficient data management system utilising MySQL database storage for recording periodic spectra data for wireless access over a private Wi-Fi network. A trapezoidal peak shaping algorithm has been implemented on an FPGA for the continuous conversation of pulses from the SiPM, signifying the detection of a radiological particle, into spectra. This system has been designed to fit within a 46 mm cylindrical diameter for in situ characterization, and can be attached to one or more SiPMs used in conjunction with a range of scintillators. LED blink tests have been used to optimise the trapezoidal shaper coefficients to maximise the resolution of the recorded spectra. Experiments with an array of SiPMs integrated with a NaI(Tl) scintillator exposed to sealed sources of Co-60, Cs-137, Na-22 and Am-241 have shown that the detector achieves a peak efficiency of 27.09 ± 0.13% for a gamma peak at 59.54 keV produced by Am-241, and a minimum energy resolution (Delta E/E) of 4.27 ± 1.16% for the 1332.5 keV gamma peak from Co-60.

AB - A compact back-end interface for silicon photomultipliers (SiPMs) implementing Zener diode-based temperature compensation has been developed for the remote detection of beta and gamma radiation. Remote detection is facilitated by the development of an efficient data management system utilising MySQL database storage for recording periodic spectra data for wireless access over a private Wi-Fi network. A trapezoidal peak shaping algorithm has been implemented on an FPGA for the continuous conversation of pulses from the SiPM, signifying the detection of a radiological particle, into spectra. This system has been designed to fit within a 46 mm cylindrical diameter for in situ characterization, and can be attached to one or more SiPMs used in conjunction with a range of scintillators. LED blink tests have been used to optimise the trapezoidal shaper coefficients to maximise the resolution of the recorded spectra. Experiments with an array of SiPMs integrated with a NaI(Tl) scintillator exposed to sealed sources of Co-60, Cs-137, Na-22 and Am-241 have shown that the detector achieves a peak efficiency of 27.09 ± 0.13% for a gamma peak at 59.54 keV produced by Am-241, and a minimum energy resolution (Delta E/E) of 4.27 ± 1.16% for the 1332.5 keV gamma peak from Co-60.

KW - silicon photomultiplier

KW - gamma detectors

KW - temperature compensation

KW - wireless data acquisition systems

U2 - 10.3390/s23084053

DO - 10.3390/s23084053

M3 - Journal article

VL - 23

JO - Sensors

JF - Sensors

SN - 1424-8220

IS - 8

M1 - 4053

ER -