Home > Research > Publications & Outputs > Evaluation of an electrochemical sensor for mea...
View graph of relations

Evaluation of an electrochemical sensor for measuring blood ketones.

Research output: Contribution to Journal/MagazineJournal articlepeer-review

Published

Standard

Evaluation of an electrochemical sensor for measuring blood ketones. / Bryne, H. A.; Tieszen, K. L.; Hollis, S. et al.
In: Diabetes Care, Vol. 23, No. 4, 2000, p. 500-503.

Research output: Contribution to Journal/MagazineJournal articlepeer-review

Harvard

APA

Vancouver

Bryne HA, Tieszen KL, Hollis S, Dornan TL. Evaluation of an electrochemical sensor for measuring blood ketones. Diabetes Care. 2000;23(4):500-503.

Author

Bryne, H. A. ; Tieszen, K. L. ; Hollis, S. et al. / Evaluation of an electrochemical sensor for measuring blood ketones. In: Diabetes Care. 2000 ; Vol. 23, No. 4. pp. 500-503.

Bibtex

@article{a4e17bbfb26b4334aa2eb5142032d755,
title = "Evaluation of an electrochemical sensor for measuring blood ketones.",
abstract = "OBJECTIVE: To evaluate the performance of a hand-held ketone sensor that is able to measure blood beta-hydroxybutyrate (beta-HBA) concentrations within 30 s in patients with diabetic ketoacidosis (DKA) and patients who attend a weight management clinic. RESEARCH DESIGN AND METHODS: Two groups of patients were studied: 19 patients admitted with DKA and 156 patients attending a weight management clinic. Paired capillary and venous whole blood samples were measured using the ketone sensor and also using an enzymatic laboratory reference method. RESULTS: The ketone sensor accurately measured beta-HBA concentrations in patients with DKA (limits of agreement -0.9 to + 1.0 mmol/l) or starvation-induced ketonemia (limits of agreement -0.5 to +0.5 mmol/l). CONCLUSIONS: This ketone sensor accurately measures whole blood beta-HBA concentrations within 30 s.",
author = "Bryne, {H. A.} and Tieszen, {K. L.} and S. Hollis and Dornan, {T. L.}",
year = "2000",
language = "English",
volume = "23",
pages = "500--503",
journal = "Diabetes Care",
issn = "1935-5548",
publisher = "American Diabetes Association Inc.",
number = "4",

}

RIS

TY - JOUR

T1 - Evaluation of an electrochemical sensor for measuring blood ketones.

AU - Bryne, H. A.

AU - Tieszen, K. L.

AU - Hollis, S.

AU - Dornan, T. L.

PY - 2000

Y1 - 2000

N2 - OBJECTIVE: To evaluate the performance of a hand-held ketone sensor that is able to measure blood beta-hydroxybutyrate (beta-HBA) concentrations within 30 s in patients with diabetic ketoacidosis (DKA) and patients who attend a weight management clinic. RESEARCH DESIGN AND METHODS: Two groups of patients were studied: 19 patients admitted with DKA and 156 patients attending a weight management clinic. Paired capillary and venous whole blood samples were measured using the ketone sensor and also using an enzymatic laboratory reference method. RESULTS: The ketone sensor accurately measured beta-HBA concentrations in patients with DKA (limits of agreement -0.9 to + 1.0 mmol/l) or starvation-induced ketonemia (limits of agreement -0.5 to +0.5 mmol/l). CONCLUSIONS: This ketone sensor accurately measures whole blood beta-HBA concentrations within 30 s.

AB - OBJECTIVE: To evaluate the performance of a hand-held ketone sensor that is able to measure blood beta-hydroxybutyrate (beta-HBA) concentrations within 30 s in patients with diabetic ketoacidosis (DKA) and patients who attend a weight management clinic. RESEARCH DESIGN AND METHODS: Two groups of patients were studied: 19 patients admitted with DKA and 156 patients attending a weight management clinic. Paired capillary and venous whole blood samples were measured using the ketone sensor and also using an enzymatic laboratory reference method. RESULTS: The ketone sensor accurately measured beta-HBA concentrations in patients with DKA (limits of agreement -0.9 to + 1.0 mmol/l) or starvation-induced ketonemia (limits of agreement -0.5 to +0.5 mmol/l). CONCLUSIONS: This ketone sensor accurately measures whole blood beta-HBA concentrations within 30 s.

M3 - Journal article

VL - 23

SP - 500

EP - 503

JO - Diabetes Care

JF - Diabetes Care

SN - 1935-5548

IS - 4

ER -