N2 - The test of three biopolymers were conducted using a modified pin-on-plate rig to evaluate whether the machine could be validated to the ASTM standard. The modification entailed the addition of rotational motion of the test pins, to give multi-directional motion. Pins were machined to form flat ended cylinders and articulated against stainless steel plates. The results show that the biopolymers subjected to multi-directional motion exhibited relevant values of wear and ASTM recommended targets were achieved.
AB - The test of three biopolymers were conducted using a modified pin-on-plate rig to evaluate whether the machine could be validated to the ASTM standard. The modification entailed the addition of rotational motion of the test pins, to give multi-directional motion. Pins were machined to form flat ended cylinders and articulated against stainless steel plates. The results show that the biopolymers subjected to multi-directional motion exhibited relevant values of wear and ASTM recommended targets were achieved.
M3 - Conference contribution/Paper
AN - SCOPUS:13844276103
SN - 1877040193
SN - 9781877040191
T3 - Transactions - 7th World Biomaterials Congress
BT - Transactions - 7th World Biomaterials Congress
T2 - Transactions - 7th World Biomaterials Congress
Y2 - 17 May 2004 through 21 May 2004
ER -