Home > Research > Publications & Outputs > Polymeric biomaterial wear test rig validated t...
View graph of relations

Polymeric biomaterial wear test rig validated to ASTM F732-00 and against clinical data

Research output: Contribution in Book/Report/Proceedings - With ISBN/ISSNConference contribution/Paperpeer-review

Published

Standard

Polymeric biomaterial wear test rig validated to ASTM F732-00 and against clinical data. / Joyce, T. J.; Monk, D.; Thompson, P. et al.
Transactions - 7th World Biomaterials Congress. 2004. (Transactions - 7th World Biomaterials Congress).

Research output: Contribution in Book/Report/Proceedings - With ISBN/ISSNConference contribution/Paperpeer-review

Harvard

Joyce, TJ, Monk, D, Thompson, P, Chiu, P, Unsworth, A & Green, SM 2004, Polymeric biomaterial wear test rig validated to ASTM F732-00 and against clinical data. in Transactions - 7th World Biomaterials Congress. Transactions - 7th World Biomaterials Congress, Transactions - 7th World Biomaterials Congress, Sydney, Australia, 17/05/04.

APA

Joyce, T. J., Monk, D., Thompson, P., Chiu, P., Unsworth, A., & Green, S. M. (2004). Polymeric biomaterial wear test rig validated to ASTM F732-00 and against clinical data. In Transactions - 7th World Biomaterials Congress (Transactions - 7th World Biomaterials Congress).

Vancouver

Joyce TJ, Monk D, Thompson P, Chiu P, Unsworth A, Green SM. Polymeric biomaterial wear test rig validated to ASTM F732-00 and against clinical data. In Transactions - 7th World Biomaterials Congress. 2004. (Transactions - 7th World Biomaterials Congress).

Author

Joyce, T. J. ; Monk, D. ; Thompson, P. et al. / Polymeric biomaterial wear test rig validated to ASTM F732-00 and against clinical data. Transactions - 7th World Biomaterials Congress. 2004. (Transactions - 7th World Biomaterials Congress).

Bibtex

@inproceedings{8aee25a72df143f6bba3c31a2af75b73,
title = "Polymeric biomaterial wear test rig validated to ASTM F732-00 and against clinical data",
abstract = "The test of three biopolymers were conducted using a modified pin-on-plate rig to evaluate whether the machine could be validated to the ASTM standard. The modification entailed the addition of rotational motion of the test pins, to give multi-directional motion. Pins were machined to form flat ended cylinders and articulated against stainless steel plates. The results show that the biopolymers subjected to multi-directional motion exhibited relevant values of wear and ASTM recommended targets were achieved.",
author = "Joyce, {T. J.} and D. Monk and P. Thompson and P. Chiu and A. Unsworth and Green, {S. M.}",
year = "2004",
month = dec,
day = "1",
language = "English",
isbn = "1877040193",
series = "Transactions - 7th World Biomaterials Congress",
booktitle = "Transactions - 7th World Biomaterials Congress",
note = "Transactions - 7th World Biomaterials Congress ; Conference date: 17-05-2004 Through 21-05-2004",

}

RIS

TY - GEN

T1 - Polymeric biomaterial wear test rig validated to ASTM F732-00 and against clinical data

AU - Joyce, T. J.

AU - Monk, D.

AU - Thompson, P.

AU - Chiu, P.

AU - Unsworth, A.

AU - Green, S. M.

PY - 2004/12/1

Y1 - 2004/12/1

N2 - The test of three biopolymers were conducted using a modified pin-on-plate rig to evaluate whether the machine could be validated to the ASTM standard. The modification entailed the addition of rotational motion of the test pins, to give multi-directional motion. Pins were machined to form flat ended cylinders and articulated against stainless steel plates. The results show that the biopolymers subjected to multi-directional motion exhibited relevant values of wear and ASTM recommended targets were achieved.

AB - The test of three biopolymers were conducted using a modified pin-on-plate rig to evaluate whether the machine could be validated to the ASTM standard. The modification entailed the addition of rotational motion of the test pins, to give multi-directional motion. Pins were machined to form flat ended cylinders and articulated against stainless steel plates. The results show that the biopolymers subjected to multi-directional motion exhibited relevant values of wear and ASTM recommended targets were achieved.