12,000

We have over 12,000 students, from over 100 countries, within one of the safest campuses in the UK

93%

93% of Lancaster students go into work or further study within six months of graduating

Home > Research > Publications & Outputs > Simple design for dilution refrigerator with ba...
View graph of relations

« Back

Simple design for dilution refrigerator with base temperature of 2.3 mK

Research output: Contribution to journalJournal article

Published

Journal publication date1/01/1994
JournalCryogenics
Journal number6
Volume34
Number of pages2
Pages549-550
Original languageEnglish

Abstract

A simple design is presented for a dilution refrigerator. Construction is not difficult. This refrigerator has been designed to allow measurements of the properties of small samples at low temperatures and to precool small nuclear demagnetization stages. The design consists of a concentric tube heat exchanger and several sintered silver discrete heat exchangers. The minimum base temperature was measured to be 2.3 mK.